242

Database

With this button the recalculation of the selected determination(s) can be
triggered with the modifications to the variables, methods, statistics or

curve evaluation made in the subwindow

Modifications

. During the recalcu-

lation a progress bar appears; the recalculation can be canceled with
alongside the bar. The results of this recalculation are entered automatically
in the subwindow

Result view

. This button is inactive until modifications

have actually been made. After the recalculation further data can be modi-
fied and then recalculation triggered again.

With this button all the modifications that have been made in reprocessing
and that have not been saved can be rejected; the original data and results
are then available again. This button is inactive until modifications have ac-
tually been made or when the recalculation has not yet been triggered.

With this button each determination that has been modified by reprocess-
ing will be saved as a new version with a version number increased by

+1

and the subwindow

Reprocessing

will be closed. This button is inactive when

the recalculation has not yet been triggered and if not all the selected de-
terminations could be reprocessed.

With this button the result of the reprocessing carried out since the last data
saving will not be saved and the subwindow

Reprocessing

will be closed.

The same function can also be triggered with in the right top corner of

the window.

Reprocessing rules

The following rules apply for reprocessing the selected and modified determina-
tions:

Number of determinations
No more than 100 determinations (including statistically linked determina-
tions) may be selected for reprocessing.

Statistically linked determinations
If determinations are selected for reprocessing that are statistically linked

with other determinations that have not been selected then these determi-
nations will also be automatically reprocessed and then saved as new ver-
sions. Determinations that are statistically linked with each other must al-
ways be reprocessed in the same sequence as they were recorded. If no
modification is made to a determination then this determination will retain

its original determination status (i.e.

original

for non-modified determina-

tions).

Start test
During the start test all device tests and device monitoring will be ignored.

Special tracks
During reprocessing the series start, series end and error tracks are not run

through again.

Tracks with "Return immediately"
If the method contains tracks for which the option

Return immediately

has

been switched on then a warning will appear to say that it can no longer be