Metrohm tiamo 1.3 Manual User Manual
Page 585
Commands
559
Lower hysteresis
0.0 ... [ 0.2 ] ... 150.0 °C
(only Titrando)
The lower hysteresis represents a tolerance range for the lower limit of the
temperature. When the temperature exceeds the lower limit plus this hys-
teresis value, the action
Temperature lower limit ok
is triggered
Upper limit
-20.0 ... [ 150.0 ] °C
(Titrando)
-170.0 ... [ 500.0 ] °C
(Titrino)
Upper limit of the temperature. When the temperature exceeds this value,
the action
Temperature upper limit exceeded
is triggered.
Upper hysteresis
0.0 ... [ 0.2 ] ... 150.0 °C
(only Titrando)
The upper hysteresis represents a tolerance range for the upper limit of the
temperature. When the temperature falls below the upper limit plus this hys-
teresis value, the action
Temperature upperlimit ok
is triggered
Action
Definition of the action to be taken when the limits of the temperature are in-
fringed:
Stop determination
(only Titrando)
The running
DOS
command is quit, then the exit track (if present) is started
and the determination is stopped.
Stop command
The running
DOS
command is quit, then the next command is started.
Wait for [Continue]
Reagent dosing in the current
DOS
command is interrupted and a message
appears. As soon as the monitored temperature is again within the limits
(including hysteresis) reagent dosing can be resumed by pressing the
[Continue]
button in this message window.
Wait for limit ok
Reagent dosing in the current
DOS
command is interrupted. As soon as the
monitored temperature is again within the limits (including hysteresis) re-
agent dosing will be resumed automatically.
[ none ]
No action will be taken.
Track call on limit exceeding
In this table, which cannot be edited directly, max. 20 entries can be de-
fined for the track that is to be started automatically when a particular limit
is infringed.
Open the dialog window
Track call #
(see below) to enter parameters of a
new track call.
Open the dialog window
Track call #
(see below) to edit parameters of the
selected track call.
Delete the selected track call.