Metrohm tiamo 1.3 Manual User Manual
Page 806
780
Configuration
Dosing unit
Hardware
Name
24 characters, [ 'empty' ]
Freely definable name for the dosing unit.
Type
Shows the type of dosing unit:
DU
Non-intelligent dosing unit without data chip.
IDU
Intelligent 807 Dosing Unit with data chip.
Order number
24 characters, [ 'empty' ]
Order number of dosing unit. With intelligent dosing units the order number
is read off automatically and cannot be edited.
Serial number
10 characters, [ 'empty' ]
(DU)
Serial number of dosing unit. With intelligent dosing units the serial number
is read off automatically and cannot be edited.
Cylinder volume
2, 5, 10, [ 20 ], 50 mL
Cylinder volume of dosing unit. With intelligent dosing units the cylinder
volume is read off automatically and cannot be edited. If in the method you
have selected a solution in a titration or dosing command then the cylinder
volume will be checked in the sequence.
Cylinder serial number
8 characters, [ 'empty' ]
Serial number of cylinder. It is printed on new cylinders. With intelligent dos-
ing units it is read off automatically. It can be altered at any time, e.g. if the
cylinder has to be replaced.