Commands

525

25 characters, [ STDADD man # ]

Name of the command.

Number of additions

1 ... [ 3 ] ... 19

Number of times the standard solution is to be added.

Volume auxiliary solution

[ 0.000 ] ... 9999.999 mL

Volume of the auxiliary solution (e.g. buffer) to be added to the sample be-
fore the first standard addition.

Addition mode

Fixed volume increments

[ on ], off

If this option is enabled a fixed volume increment muss be added for each
standard addition.

Addition volume

[ 0.00000 ] ... [ 0.100 ] ... 99999.9 mL

Quantity of the fixed volume increment.

Variable volume increments

on, [ off ]

If this option is enabled the defined volume increment muss be added for

each standard addition.

Addition volume 1...19

[ 0.00000 ] ... [ 0.100 ] ... 99999.9 mL

Quantity of the defined volume increments.

Addition solution

Concentration

0.001 ...[ 1.000 ] ... 999999.999

Concentration of addition solution.

Unit

mol/L, mmol/L, µmol/L, [ ppm ], g/L, mg/L, µg/L, mg/mL, µg/mL, %, mEq/L

Concentration unit of addition solution.