Series end track, Exit track – Metrohm tiamo 1.3 Manual User Manual
Page 338

312
Method
Series end track
An series end track is a track that can be created manually with
Insert, New
track, Series end track
or with the icon . It starts with a start command
SERIES
END
and finishes with a end command
END
. A method can only contain one series
end track.
The series end track is executed only once at the end of the last determination of
a series after the Main track. Sample Processors for example can be put in the
desired condition after a determination series.
Exit track
An exit track is a track that can be created manually with
Insert, New track, Exit
track
or with the icon . It starts with a start command
EXIT
and finishes with a
end command
END
. A method can only contain one exit track.
The exit track is executed as soon as all Normal tracks and the Main track are fin-
ished or if the determination has been stopped manually. If a new exit trackis cre-
ated a
DATABASE
and a
REPORT
command are automatically inserted in this track.
With these commands it can be guaranteed that a data base entry and a report
are created for each determination even if the determination has been stopped ei-
ther manually or due to an error.