Meas conc – Metrohm tiamo 1.3 Manual User Manual
Page 540
514
Method
Minimum evaluation Temperature
on, [ off ]
For the minimum measured value the associated time and temperature are
interpolated from the list of measuring points) (see Titration commands -
Evaluation - Minimum/Maximum evaluation).
Threshold value
0.1 ... [ 1.0 ] ... 20.0 °C/s
The evaluation of the minimum begins as soon as the slope of the curve
exceeds the threshold value.
Maximum evaluation Temperature
on, [ off ]
For the maximum measured value the associated time and temperature are
interpolated from the list of measuring points) (see Titration commands -
Evaluation - Minimum/Maximum evaluation).
Threshold value
0.1 ... [ 1.0 ] ... 20.0 °C/s
The evaluation of the maximum begins as soon as the slope of the curve
exceeds the threshold value.
Additional measured values
Screenshot and parameters: see DET pH - Additional measured values
MEAS Conc
Command for concentration measurements (direct measurement).
Devices
This command can be executed with the following devices:
Titrando: 808, 809, 835, 836, 855, 857
Appearance
The command has the following appearance:
Parameters
The parameters for the command
MEAS Conc
are configured on the following 3
tabs:
•
General/Hardware
Parameters for device, sensor and stirrer.
•
Measuring parameters
Parameters for the measurement.
•
Additional measured values
Defining further measured values which can be saved as additional col-
umns in the list of measuring points.