Commands
527
Temperature
Temperature
-20.0 ... [ 25.0 ] ... 150 °C
Manually entered measuring temperature. If a temperature sensor is con-
nected and
temperature measurement
on the tab
General/Hardware
is set to
automatic
or
continuous
, the temperature is measured continuously. The
value is used for the temperature adjustment during the measurement.
Switch off stirrer during measurement
on, [ off ]
If this option is switched on then the stirrer is switched off during the meas-
urement. In this case the two following parameters can be edited:
Stir before measurement
[ 0 ] ... 999999 s
Stirring time between end of dosing and measurement.
Pause before measurement
[ 0 ] ... 999999 s
Waiting time before the measurement during which no stirring takes place.
STDADD dos
Command for Standard addition with addition of the standard addition solution
from a dosing device.
Devices
This command can be executed with the following devices:
Titrando: 808, 809, 835, 836, 855, 857
Appearance
The command has the following appearance:
Parameters
The parameters for the command
STDADD dos
are configured on the following 3
tabs:
•
General/Hardware
Parameters for device, dosing device, sensor and stirrer.
•
Standard addition
Parameters for standard addition.
•
Measuring parameters
Parameters for the measurement.
Standard addition curve
The standard addition curve with its evaluation data for a determination is shown
in the program part Database in the dialog window
Calibration curve
which is
opened with the menu item
Determinations, Show calibration curve...
or the symbol
.