Metrohm tiamo 1.3 Manual User Manual
Page 840
814
Configuration
Beaker radius
Radius of the special beaker.
Beaker sensor
Shows whether and which beaker sensor is to be used for the special
beaker.
Opens the dialog window
Special beaker #
for editing the data of the se-
lected special beaker.
Special beaker
In the dialog window
Special beaker #
the parameters for the special beaker se-
lected in the Table of special beakers can be defined.
Rack position
[ 0 ] ... n (n is rack-dependent)
Number of rack position for selected special beaker.
Note
Special beakers should preferably be set at high rack positions, so that a sam-
ple series can start at rack position 1. Rack positions that are defined as special
beakers are no longer available as sample positions. They will be jumped over
during an automatic movement to a sample position.
Work position Tower 1
[ 0 ] ... 235 mm
Work position of the selected special beaker at Tower 1.
Work position Tower 2
[ 0 ] ... 235 mm
Work position of the selected special beaker at Tower 2.
Beaker radius
1.0 ... 100.0 mm, [ off ]
Radius of the selected special beaker on the rack. If the lift is moved to the
work position then this value will be compared with the parameter
Min.
beaker radius
(see Properties Tower #) that can be specifically defined for
each tower. If Beaker radius samples < Min. beaker radius then a cor-
responding error message will appear. With
off
the beaker radius will not be
checked.
Beaker sensor
[ Tower ], Robotic arm, off
When the selected special beaker is moved to with the command
MOVE
then the beaker sensor (
Tower
,
Robotic arm
) checks whether the special
beaker is present or not. With
off
no check will be made. For the option
Ro-
botic arm
a swing head with beaker sensor must be installed and a suitable
work position with beaker contact must be defined for the lift, as this must
move to the beaker recognition position.
With the parameter
Beaker test
in the command
MOVE
you can determine
whether the determination is to be terminated if a beaker is missing with or
without the display of a corresponding message and whether the series is
to be continued or also terminated.