Commands

555

Command name

25 characters, [ DOS pH # ]

Name of the command.

Dosing criterion

[ Volume/Dosing rate ], Volume/Dosing time, Dosing rate/Dosing time

Selects the two parameters that can be preset for dosing. The third pa-

rameter is calculated automatically from the formula

Volume = dosing time *

dosing rate

and is not shownDOS - monitored dosing.

Volume

0.00000 ... [ 10.0000 ] ... 99999.9 mL

Fixed volume to be added. This parameters is only displayed for

Dosing cri-

terion =

Volume/Dosing rate

or

Volume/Dosing time.

Dosing rate

[ maximum ]
0.01 ... 166.00 mL/min

(Titrando)

0.01 ... 150.00 mL/min

(Titrino)

Speed at which dosing is carried out. The maximum dosing rate depends
on the cylinder volume of the exchange unit or dosing unit used. In case the

entered dosing rate is too high for the selected dosing device, the rate will
be reduced automatically to the highest possible value. This parameters is
only displayed for

Dosing criterion =

Volume/Dosing rate

or

Volume/Dosing

time.

Note
Lower dosing rates should be entered for solutions with higher viscosities.

Dosing time

0 ... [ 100 ] ... 9999999 s

Time during which dosing is to be carried out. This parameter is only dis-
played for

Dosing criterion =

Volume/Dosing time

or

Dosing rate/Dosing time..

Pause

[ 0 ] ... 999999 s

Waiting period before dosing is started.

Temperature

-20.0 ... [ 25.0 ] ... 150 °C

(Titrando)

-170.0 ... [ 25.0 ] ... 500 °C

(Titrino)