Method

Commands

Depending on the measured value one of the following two

STAT

commands can

be chosen:

STAT pH

STAT Titration with pH electrodes (measuring quantity pH).

STAT U

STAT Titration with metal electrodes (measuring quantity U).

Control range

Set a large control range for steep curves and a small control range for flat ones.
A good approximation for the start of the control range is given by the point where
the tangents intersect.

STAT pH

Command for STAT titrations with measured value pH.

Devices

This command can be executed with the following devices which have the

STAT

pH

mode:

Titrino: 718, 736, 751, 799
Titrando: 835, 836, 842, 855, 857

Appearance

The command has the following appearance:

Parameters

The parameters for the command

STAT pH

are configured in the following 8 tabs:

General/Hardware

Parameters for devices, dosing devices, sensors and stirrers.