Subwindow Sensors
791
Order number
24 characters, [ 'empty' ]
Order number of the sensor (read-only for intelligent sensors).
Sensor serial number
10 characters, [ 'empty' ]
Serial number of the sensor (read-only for intelligent sensors).
Device
Select device, [ 'empty' ]
For non-intelligent sensors it is possible to select the device to which the
sensor is connected. All the devices in the device table which have a
measuring input are shown in the selection list. The connection of a sensor
to a particular device is not checked during the run.
For intelligent sensors the device to which the sensor is connected is
shown here automatically.
Measuring input
[ 1 ], 2
For non-intelligent sensors it is possible to select the measuring input to
which the sensor is connected. The connection of a sensor to a particular
measuring input is not checked during the run. This parameter only ap-
pears if a
Device
has been chosen.
For intelligent sensors the measuring input to which the sensor is con-
nected is shown here automatically.
Comment
24 characters
It is possible to enter remarks about the sensor.
Set to work
Select date
The current date is automatically entered here for a newly added sensor.
When editing an existing sensor the date can be selected by pressing
in the dialog window
Select date
(see General program functions - Edit - Se-
lect date).
Sensor monitoring
on, [ off ]
If this option is switched on then the working life of the sensor will be moni-
tored.
Working life
0 ... [ 999 ] days
Working life of the sensor in days. If a value is entered here then the
Expiry
date
will be adapted automatically.
Expiry date
Select date, [ Set to work + 999 days ]
Expiry date of the sensor. This date can be selected by pressing
in the
dialog window
Select date
(see General program functions - Edit - Select
date). If a date is entered here then the value for the
Working life
will be
adapted automatically.
Action
If during sensor monitoring it is found that the working life has expired then one of
the following actions will be triggered automatically at Start test:
Record message
The message that the working life of the sensor has expired will be auto-
matically saved with the determination.
Display message
A message will appear and you can choose whether to continue with the
run or cancel it. If the run is continued then the message that the working