Metrohm tiamo 1.3 Manual User Manual
Page 530

504
Method
Measuring time
0 ... [ 120 ] ... 999'999 s
(Titrando)
0 ... [ 120 ] ... 9'999 s
(Titrino)
Maximum period of time for measurement.
Measuring interval
0.1 ... [ 2.0 ] ... 999'999.0 s
(Titrando, increment = 0.1 s)
0.08 ... [ 2.0 ] ... 16'200 s
(Titrino, increment = 0.08 s)
Time interval for entering a measuring point in the list of measuring points.
Stop measured value
-200.0 ... 200.0 µA, [ off ]
(Titrando only)
Stop when the preset measured value has been reached since the start of
the measurement.
off
means no stop.
Temperature
Temperature
-20.0 ... [ 25.0 ] ... 150 °C
(Titrando)
-170.0 ... [ 25.0 ] ... 500 °C
(Titrino)
Manually entered measuring temperature. If a temperature sensor is con-
nected and
temperature measurement
on the tab General/Hardware is set to
automatic
or
continuous
, the temperature is measured continuously. The
value is used for the temperature adjustment during the measurement.
Evaluations
The following methods for evaluating measuring curves are defined on this tab.
Command name
25 characters, [ MEAS Upol # ]
Name of the command.
Fix end point evaluation
on, [ off ]
To a fixed quantity (measured value or time) the associated values for the
other quantities are interpolated from the list of measuring points for the
fixed end point.