Commands
567
Upper hysteresis
0.0 ... [ 2.0 ] ... 2000.0 mV
(only Titrando)
The upper hysteresis represents a tolerance range for the upper limit of the
measured value. When the measured value falls below the upper limit plus
this hysteresis value, the action
Measured value upper limit ok
is triggered.
Action
D
efinition of the action to be taken when the limits of the measured value
are infringed:
Stop determination
(only Titrando)
The running
DOS
command is quit, then the exit track (if present) is started
and the determination is stopped.
Stop command
The running
DOS
command is quit, then the next command is started.
Wait for [Continue]
Reagent dosing in the current
DOS
command is interrupted and a message
appears. As soon as the monitored measured value is again within the lim-
its (including hysteresis) reagent dosing can be resumed by pressing the
[Continue]
button in this message window.
Wait for limit ok
Reagent dosing in the current
DOS
command is interrupted. As soon as the
monitored measured value is again within the limits (including hysteresis)
reagent dosing is continued automatically.
[ none ]
No action will be taken.
Monitoring Temperature
on, [ off ]
If this option is switched on then the temperature will be monitored and any
infringements will be entered in the list of measuring points.
Lower limit
[ -20.0 ] ... 150.0 °C
(Titrando)
[ -170.0 ] ... 500.0 °C
(Titrino)
Lower limit of the temperature. When the temperature falls below this value,
the action
Temperature lower limit exceeded
is triggered.
Lower hysteresis
0.0 ... [ 0.2 ] ... 150.0 °C
(only Titrando)
The lower hysteresis represents a tolerance range for the lower limit of the
temperature. When the temperature exceeds the lower limit plus this hys-
teresis value, the action
Temperature lower limit ok
is triggered
Upper limit
-20.0 ... [ 150.0 ] °C
(Titrando)
-170.0 ... [ 500.0 ] °C
(Titrino)
Upper limit of the temperature. When the temperature exceeds this value,
the action
Temperature upper limit exceeded
is triggered.
Upper hysteresis
0.0 ... [ 0.2 ] ... 150.0 °C
(only Titrando)
The upper hysteresis represents a tolerance range for the upper limit of the
temperature. When the temperature falls below the upper limit plus this hys-
teresis value, the action
Temperature upper limit ok
is triggered
Action
Definition of the action to be taken when the limits of the temperature are in-
fringed:
Stop determination
(only Titrando)
The running
DOS
command is quit, then the exit track (if present) is started
and the determination is stopped.
Stop command
The running
DOS
command is quit, then the next command is started.