beautypg.com

Metrohm tiamo 1.3 Manual User Manual

Page 422

background image

396

Method

Dosing rate

0.01 ... 166.00 mL/min, [ maximum ]

(Titrando),

0.01 ... 150.00 mL/min, [ maximum ]

(Titrino)

Speed at which the start volume is to be added. The maximum dosing rate
depends on the cylinder volume of the used exchange unit or dosing unit.

Pause 2

Pause 2

[ 0 ] ... 999999 s

Waiting time, e.g. until the electrode has settled down after the start or a re-

action time after the addition of a start volume. The pause takes place at
the end of all start conditions.