beautypg.com

Metrohm tiamo 1.3 Manual User Manual

Page 444

background image

418

Method

Minimum evaluation

on, [ off ]

For the minimum measured value the associated volume, time and tem-

perature are interpolated from the list of measuring points (see Titration
commands - Evaluation - Minimum/maximum evaluation
).

Threshold value

0.5 ... [ 5.0 ] ... 10.0 µA/mL

The evaluation begins as soon as the slope of the curve exceeds the set
threshold value.

Maximum evaluation

on, [ off ]

For the maximum measured value the associated volume, time and tem-
perature are interpolated from the list of measuring points (see Titration
commands - Evaluation - Minimum/maximum evaluation
.

Threshold value

0.5 ... [ 5.0 ] ... 10.0 µA/mL

The evaluation begins as soon as the slope of the curve exceeds the set

threshold value.

Additional measured values

Screenshot and parameters: see DET pH - Additional measured values