MTS Series 793 Application User Manual
Page 78

MTS MultiPurpose TestWare®
MPT Options Editor Specimen Log Tab
MPT Options Editor
78
•
Fatal Error—Writes only Fatal Error messages; does not write
Diagnostic, Information, Warning, and Error messages.
Source—Selects the information source from which the log is built.
•
This Application Only—Records messages from the MPT application
only.
•
All Applications—Records messages from other sources, such as the
Station Manager application.
Archive Auto Deletion
Important
Enabling this control may result in the inadvertent deletion of
specimen log entries. The Specimen Log is unlikely to exceed
the internal limit of 500 Specimen Log files for a given MPT
specimen (unless your procedure is generating an excessive
number of messages). Therefore, MTS recommends that you
disable this control if you routinely use new MPT specimens from
test-to-test. Disabling this control ensures that you are able to
recover relevant log data for your MPT specimens in the future.
This control enables or disables the auto deletion function and sets the deletion
interval in days.
The Specimen Log archive has an internal limit of 500 Specimen Log files,
with each file containing approximately 1000 messages. Once 500 archived
files are accumulated and a new archive file is generated, the oldest archive file
is automatically deleted to make room for the newest archive file.
The Archive Auto Deletion control lets you program when archived files are
deleted in terms of days (within the internal limit of 500 archived files).
To enable the Archive Auto Deletion function, click in the Delete Older Than
checkbox such that a check mark appears in the box. If the check mark appears
in the box, the function is enabled.
Set the deletion interval using the scroll arrows to change the number of days
displayed. The scroll arrows are only active if the Archive Auto Deletion
function is enabled.
Specimen Log Tab (part 2 of 2)
I
TEM
D
ESCRIPTION