Auto Offset Process
About the Auto Offset Process
The Auto Offset process allows you to zero input signals that are not being used
for the active control mode. The process completes after all selected signals have
been zeroed. If a selected signal is active when the process is triggered, an error
will be logged.
Auto Offset process
example
Suppose you want to run a sweep test on a pliable specimen in force control. As
the specimen becomes more compliant, displacement will increase, so you may
want to apply an offset to displacement after each pass.
•
You can achieve this by adding a Sweep Command process and Auto
Offset process to a Group process. Configure the test so that the Auto
Offset process triggers after each pass of the Sweep Command. As
displacement increases, the signal will be continually zeroed.