beautypg.com

Rs-485 communications – WattMaster WM-WCC3-TGD-01B User Manual

Page 464

background image

10. RS-485 COMMUNICATIONS

WCC III Technical Guide

10-12

SAT 3C/D/F Troubleshooting with a Digital Voltmeter

Figure 10-13: Typical SAT 3C/D/F TUC RS-485 Satellite Communications DC Voltage Measurements with the

TUC COMM Loop Connected

Figure 10-14: Typical SAT 3C/D/F TUC RS-485 Satellite Communications DC Voltage Measurements with the

TUC COMM Loop Disconnected

The voltage measurements in Figure 10-13 are approximate voltages. These voltage measurements are taken when the power to the SAT
3C/D/F Controller is “ON” and the RS-485 TUC EXP communications loop is connected. These two voltages will fl uctuate slightly and
also will momentarily “fl ip” to the other meter reading’s value. The voltage measurement from “T” to “SHIELD” should be around 2.7
VDC. The voltage measurement from “R” to “SHIELD” should be around 3.00 VDC. Typical bad voltage measurement values would
be anything above 3.8 VDC and anything below 1.5 VDC. Helpful hint: Removal of the SAT 3C/D/F cover may aid in measuring the
communication loop voltages.

The voltage measurements in Figure 10-14 are approximate voltages. These voltage measurements are taken when the power to the SAT
3C/D/F Controller is “ON” and the RS-485 TUC EXP communications loop is disconnected. These two voltages will not fl uctuate. The
voltage measurement from “T” to “SHIELD” should be around 3.25 VDC. The voltage measurement from “R” to “SHIELD” should be
around 3.25 VDC. Typical bad voltage measurement values would be anything above 3.8 VDC and anything below 1.5 VDC. Helpful hint:
Removal of the SAT 3C/D/F cover may aid in measuring the communication loop voltages.

This manual is related to the following products: