3.3 Sample tables
■■■■■■■■■■■■■■■■■■■■■■
86
■■■■■■■■
viva 1.0
NOTE
If a value is entered or imported into this field, then the time of the
data entry and the data source (manual, 'Name of the barcode
reader' or 'Name of the import file') are additionally stored auto-
matically, placed in the determination and identified in the information
subwindow as entry date and data source.
NOTE
If a negative value is entered into this field, then this will also be calcu-
lated as a negative in the formula editor.
Entry
Number with 10 digits at most
Default value
1.0
Sample amount unit
Unit of the sample amount. Any text can be entered into this field.
Entry
16 characters
Selection
L | mL | µL | g | mg | µg | pieces
Default value
mL
Analytical volume
Volume portion (aliquot) of the diluted sample's dilution volume used for
the determination (see Glossary, page 865). This volume is added to the
measuring vessel with the ADD SAMPLE command.
Input range
-1E+99 - 1E+99 (max. 10 places) mL
Default value
'empty' mL
Dilution volume
Maximum filling volume used for dilution of the sample (see Glossary,
page 868).
Input range
-1E+99 - 1E+99 (max. 10 places) mL
Default value
'empty' mL
Navigation
Display of the currently selected line in the sample table with the following
functions: