11 calibration factor dt – Metrohm viva 1.0 Manual User Manual
Page 623
■■■■■■■■■■■■■■■■■■■■■■
5 Method
viva 1.0
■■■■■■■■
611
5.6.7.7.10
Relative standard deviation of the effective addition volume of the
sample solution for the evaluation ratio
Variable name
RS.’Command name’.’Substance name’.VEFFSTD.RSD
Description
Relative standard deviation of the effective addition volume of the added
sample solution for the evaluation ratio.
Data type
Number
Decimal places
0…1…6
Unit
%
Remark
Only for Calibration method = DT and Sample type = Sample.
5.6.7.7.11
Calibration factor DT
Variable name
RS.’Command name’.’Substance name’.CALFACT
Description
The calibration factor corresponds to the concentration of the standard in
the measuring vessel for the evaluation ratio.
Decimal places
0…2…6
Unit
#g/L, #L/L, #mol/L
The unit corresponds to the added standard solution.
Remark
Only for Calibration method = DT and Sample type = Standard.
Calculation
Variable
Description
CALFACT
Calibration factor DT.
CONC(STD)
Concentration of the standard solution for creating the
calibration curve.
VOL(STD)
Uncorrected addition volume of the standard solution
for the evaluation ratio.
VTOT(VMS,
Cal)
Total volume in the measuring vessel at the moment of
the measurement of the VMS by calling the measuring
command via the command CALL VMS at the moment
of calibration.