5 Method

viva 1.0

■■■■■■■■

611

5.6.7.7.10

Relative standard deviation of the effective addition volume of the
sample solution for the evaluation ratio

Variable name

RS.’Command name’.’Substance name’.VEFFSTD.RSD

Description

Relative standard deviation of the effective addition volume of the added
sample solution for the evaluation ratio.

Data type

Number

Decimal places

0…1…6

Unit

%

Remark

Only for Calibration method = DT and Sample type = Sample.

5.6.7.7.11

Calibration factor DT

Variable name

RS.’Command name’.’Substance name’.CALFACT

Description

The calibration factor corresponds to the concentration of the standard in
the measuring vessel for the evaluation ratio.

Decimal places

0…2…6

Unit

#g/L, #L/L, #mol/L

The unit corresponds to the added standard solution.

Remark

Only for Calibration method = DT and Sample type = Standard.

Calculation

Variable

Description

CALFACT

Calibration factor DT.

CONC(STD)

Concentration of the standard solution for creating the
calibration curve.

VOL(STD)

Uncorrected addition volume of the standard solution
for the evaluation ratio.

VTOT(VMS,
Cal)

Total volume in the measuring vessel at the moment of
the measurement of the VMS by calling the measuring
command via the command CALL VMS at the moment
of calibration.