beautypg.com

2 editing templates for output lines – Metrohm viva 1.0 Manual User Manual

Page 664

background image

6.3 Configuration data

■■■■■■■■■■■■■■■■■■■■■■

652

■■■■■■■■

viva 1.0

The [Edit] menu below the list of templates or a right-click on the table
opens a menu with the following menu items:

New...

Create a new template (see Chapter 6.3.3.2.2, page 652).

Properties...

Edit the selected template (see Chapter 6.3.3.2.2, page 652).

Delete

Deletes the selected template.

6.3.3.2.2

Editing templates for output lines

Dialog window: Configuration

Tools Templates Output lines... Templates

for output lines

New... / Properties... New template / Edit template

Signal name

Name of the template for the output signal.

Entry

25 characters

Output signal

Entry of the bit pattern for the output signal with exactly 14 characters.

It is possible to enter the characters

0 = Line inactive,

1 = Line active,

* = Any line status and

p = Set pulse

Selection

Bit pattern with exactly 14 characters (0, 1, *,
p)
| **************

Default value

**************

The output lines and bits are numbered from right to left:

Output 13 12 11 10 9 8 7 6 5 4 3 2 1 0

Bit

13 12 11 10 9 8 7 6 5 4 3 2 1 0

NOTE

We recommend masking the irrelevant output lines with an asterisk *
so as not to modify these line conditions.

Pulse length

Duration of the transmitted pulse.

Input range

1 - 999 s

Default value

1 s