4.5 Determination overview

■■■■■■■■■■■■■■■■■■■■■■

264

■■■■■■■■

viva 1.0

This button is disabled for as long as recalculation has not yet been trig-
gered and if not all the selected determinations were able to be recalcula-
ted.

[Cancel] or

Closes the Reprocess dialog window. The result of the reprocessing
made since the last saving will not be saved.

4.5.2.5.3

Recalculation rules

Dialog window: Reprocess

The following rules apply for recalculating the selected and modified
determinations:

Number of determinations
No more than 100 determinations may be selected for recalculation.

Start test
During the start test all device tests and device monitoring will be
ignored.

Special tracks
Series start track, series end track and error tracks are not run through
during recalculation.

Tracks with "Return immediately"
If the method contains tracks with the option Return immediately
enabled, then a warning will will appear, because it can no longer be
guaranteed that during recalculation the method will run in the same
way as during the determination.

Monitoring If the method contains monitoring of variables and
results, then the corresponding messages will be shown and included
in the report, but no e-mails will be sent and no beeps produced. If
one of the two actions Cancel determination or Cancel determi-
nation and series
is activated, recalculation is continued neverthe-
less.

Measuring commands
For the commands CVS and CAL MEAS Opt, only the evaluations
with existing data will be recalculated; all other parameters will be
ignored.

Conditional CALL commands
In the case of a conditional CALL command, the track that was run
through originally will always be called.

Waiting times and timeout times
All waiting times and timeout times of commands are set to 0 for
recalculation.