4.5 Determination overview
■■■■■■■■■■■■■■■■■■■■■■
264
■■■■■■■■
viva 1.0
This button is disabled for as long as recalculation has not yet been trig-
gered and if not all the selected determinations were able to be recalcula-
ted.
[Cancel] or
Closes the Reprocess dialog window. The result of the reprocessing
made since the last saving will not be saved.
4.5.2.5.3
Recalculation rules
Dialog window: Reprocess
The following rules apply for recalculating the selected and modified
determinations:
■
Number of determinations
No more than 100 determinations may be selected for recalculation.
■
Start test
During the start test all device tests and device monitoring will be
ignored.
■
Special tracks
Series start track, series end track and error tracks are not run through
during recalculation.
■
Tracks with "Return immediately"
If the method contains tracks with the option Return immediately
enabled, then a warning will will appear, because it can no longer be
guaranteed that during recalculation the method will run in the same
way as during the determination.
■
Monitoring If the method contains monitoring of variables and
results, then the corresponding messages will be shown and included
in the report, but no e-mails will be sent and no beeps produced. If
one of the two actions Cancel determination or Cancel determi-
nation and series is activated, recalculation is continued neverthe-
less.
■
Measuring commands
For the commands CVS and CAL MEAS Opt, only the evaluations
with existing data will be recalculated; all other parameters will be
ignored.
■
Conditional CALL commands
In the case of a conditional CALL command, the track that was run
through originally will always be called.
■
Waiting times and timeout times
All waiting times and timeout times of commands are set to 0 for
recalculation.