Metrohm viva 1.0 Manual User Manual
Page 766
7.4 919 IC Autosampler plus
■■■■■■■■■■■■■■■■■■■■■■
754
■■■■■■■■
viva 1.0
Move to initial position
on | off (Default value: off)
If this check box is activated, the position defined here is approached
when the device is initialized.
Tower
Selection of the tower on the sample changer for moving to the required
position.
Only editable for Move to initial position = on and if the corresponding
sample changer has two towers, otherwise inactive.
Selection
1 | 2
Default value
1
Move
Selection of the target position which is to be approached:
Only editable for Move to initial position = on, otherwise inactive.
Selection
Rack position | Special beaker | External posi-
tion
Default value
Rack position
Rack position
Predefined position on the rack. The lift is lowered to the work position
defined for that position.
Special beaker
Special position on the rack defined in the rack properties. The lift is
lowered to the work position defined for that position.
External position
External position which is defined for the Swing Head mounted to the
tower. The lift is lowered to the work position defined for that posi-
tion.
Number
Specification of the number of the rack position, of the special beaker or
of the external position.
Only editable for Move to initial position = on, otherwise inactive.
for Move = Rack position
Input range
1 - 999
Default value
1
for Move = Special beaker
Input range
1 - 16
Default value
1