5.5 Method run subwindow

■■■■■■■■■■■■■■■■■■■■■■

486

■■■■■■■■

viva 1.0

'Name of the dosing unit'
Alphabetically organized selection of the dosing units from the table of
dosing units in the configuration.
not defined
The dosing unit must be assigned with not defined by the user at the
start of the method.

Dosing rate

Rate (volume per time unit) of expulsion. The maximum rate depends on
the cylinder volume of the buret used and on the diameter of the hose at
the selected dosing port.

Input range

0.01 - 166.00 mL/min

Default value

2.00 mL/min

Selection

maximum

CAUTION

The maximum permitted dosing rate of 4 mL/min may not be exceeded
for capillary tubes with a diameter < 1 mm.

Filling rate

Rate (volume per time unit) of filling. The maximum rate depends on the
cylinder volume of the buret used and on the diameter of the hose at the
selected fill port.

Input range

0.01 - 166.00 mL/min

Selection

maximum

Default value

maximum

5.5.2.7.6

LQH

5.5.2.7.6.1

LQH - Overview

Dialog window: Method

LQH Properties... LQH - 'Command name'

Command for complex dosing tasks with a dosing device of the Dosino
800 type.

NOTE

Volumes dosed into the measuring cell with the LQH command are not
taken into account in the volume bookkeeping of a quantitative deter-
mination. It is mandatory that the ADD commands be used in such
cases.