2 loop - properties – Metrohm viva 1.0 Manual User Manual
Page 559
![background image](/manuals/586912/559/background.png)
■■■■■■■■■■■■■■■■■■■■■■
5 Method
viva 1.0
■■■■■■■■
547
5.5.2.10.3.2
LOOP - Properties
Dialog window: Method
▶ LOOP ▶ Properties... ▶ LOOP - 'Command name'
Command name
Name of the command.
Entry
25 characters
Stop criteria
Definition of three possible stop criteria. The stop criterion that is reached
first leads to the cancellation of the loop.
Max. run number
on | off (Default value: on)
If this check box is activated, the maximum number of runs is used as a
stop criterion for repeat and while loops. If the maximum number of runs
has been reached, the loop will be canceled the next time the loop com-
mand is called.
Input range
0 - 999
Default value
2
Maximum run time
on | off (Default value: off)
If this check box is activated, the maximum run time is used as a stop cri-
terion for repeat and while loops. The starting point for the run time is the
first start of the loop command, i.e. the run time for while loops does not
start until after the first loop run has been started. If the maximum run
time has been reached, the loop will be canceled the next time the loop
command is called.
Input range
0 - 999.9
Default value
0
Selection
min | s
Default value
min
NOTE
The run time is also interrupted if the method is stopped with [HOLD].