5.6 Evaluation subwindow
■■■■■■■■■■■■■■■■■■■■■■
608
■■■■■■■■
viva 1.0
Variable
Description
HGTNORM
Standardized height.
HGT
Peak height = RS.’Command name'.'Substance
name'.VAR{x}.REP{y}.HGT.
HGT.MNV
Mean value of the peak heights of all replications for
VMS = RS.’Command name'.'Substance
name'.VMS.REP{y}.HGT.MNV (for Calibration
method = DT).
5.6.7.7.3
Effective addition volume of the standard solution for the evalua-
tion ratio
Variable name
RS.’Command name’.’Substance name’.VEFFSTD
Description
Effective addition volume of the added standard solution for the evalua-
tion ratio (see Glossary, page 868).
Decimal places
0…2…6
Unit
#L
Remark
Only for Calibration method = DT and Sample type = Standard. This
value is required as x value for the creation of the calibration curve.
Calculation
Variable
Description
VEFFSTD
Effective addition volume of the added standard solution
for the evaluation ratio.
VOL(STD)
Uncorrected addition volume of the standard solution
for the evaluation ratio.
VTOT(VMS,
Cal)
Total volume in the measuring vessel at the moment of
the measurement of the VMS by calling the measuring
command via the command CALL VMS at the moment
of calibration.