5 Method
viva 1.0
■■■■■■■■
409
Cleaning
Cleaning potential
Potential that is applied to the electrodes during the Cleaning time. This
potential can be used e.g. for the electrochemical cleaning of solid-state
electrode surfaces that have been contaminated with products of elec-
trode redox processes.
Input range
-5.000 - 5.000 V (Increment: 0.001)
Selection
off
Default value
off
Cleaning time
Waiting time during which the Cleaning potential is applied to the elec-
trodes.
Only editable for Cleaning potential
≠ off, otherwise inactive.
Input range
0.0 - 9,999.9 s (Increment: 0.1)
Default value
0.0 s
Standby potential
Standby potential
The standby potential is applied to the electrodes after completion of the
sweep. It remains in effect until it is switched off manually or until a new
voltage is applied to the electrodes. If no standby potential is defined,
then the voltage applied to the electrodes is automatically switched off
after the completion of the sweep.
Input range
-5.000 - 5.000 V (Increment: 0.001)
Selection
off
Default value
off
5.5.2.3.2.6
CVS - Potentiostat
Tab: Method
▶ CVS ▶ Properties... ▶ Potentiostat
Command name
Name of the command.
Entry
25 characters
Current measuring range
Maximum range
Limitation of the highest current measurement range.