6 cvs - potentiostat – Metrohm viva 1.0 Manual User Manual

Page 421

background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

viva 1.0

■■■■■■■■

409

Cleaning

Cleaning potential

Potential that is applied to the electrodes during the Cleaning time. This
potential can be used e.g. for the electrochemical cleaning of solid-state
electrode surfaces that have been contaminated with products of elec-
trode redox processes.

Input range

-5.000 - 5.000 V (Increment: 0.001)

Selection

off

Default value

off

Cleaning time

Waiting time during which the Cleaning potential is applied to the elec-
trodes.

Only editable for Cleaning potential

off, otherwise inactive.

Input range

0.0 - 9,999.9 s (Increment: 0.1)

Default value

0.0 s

Standby potential

Standby potential

The standby potential is applied to the electrodes after completion of the
sweep. It remains in effect until it is switched off manually or until a new
voltage is applied to the electrodes. If no standby potential is defined,
then the voltage applied to the electrodes is automatically switched off
after the completion of the sweep.

Input range

-5.000 - 5.000 V (Increment: 0.001)

Selection

off

Default value

off

5.5.2.3.2.6

CVS - Potentiostat

Tab: Method

CVS Properties... Potentiostat

Command name

Name of the command.

Entry

25 characters

Current measuring range

Maximum range

Limitation of the highest current measurement range.