5 Method

viva 1.0

■■■■■■■■

445

Device type

Display or selection of the instrument type. If an instrument is selected in
the Device name field, the Device type field can no longer be edited
and the instrument type belonging to the instrument is displayed.

If the not defined option is selected in the Device name field, all instru-
ment types or groups with which the command can be executed can be
selected, regardless of the instruments present in the device table.

Selection

Avantes Spectrometer | 089 Photometer

Default value

Avantes Spectrometer

Sensor

Colorimetric sensor

Selection of a colorimetric sensor from the table of colorimetric sensors.
The calibration data for the colorimetric sensor will be adopted for the
determination.

Selection

'Name of the colorimetric sensor' | empty or
'first list entry'

Default value

empty or 'first list entry'

Temperature measurement

Type of temperature measurement.

Displayed only when Device type = 089 Photometer.

Selection

continuous | off

Default value

continuous

continuous
A temperature sensor must be connected. The temperature will then
be measured continuously.
off
The temperature will not be measured. The temperature entered man-
ually under Temperature on the Measuring parameters tab will be
used.

Stirrer
This section is displayed only when Device type = 089 Photometer.

Stirrer

Selection of the stirrer.

Selection

1 | off

Default value

1

off
Means that no stirrer will be used.