8.4 Manual control - Functions
■■■■■■■■■■■■■■■■■■■■■■
808
■■■■■■■■
viva 1.0
Moves to current rack position –1.
Moves to current rack position +1.
Lift position
Set the lift position for the selected tower.
Current position
Shows the current lift position in mm.
Target position
Selection or entry of the lift position that is to be approached for the cur-
rent rack position (regular beaker, special beaker, external position).
Input range
0 - max. stroke path defined (235 mm) mm
Default value
126 mm
Selection
Home position | Work position | Shift posi-
tion | Rinse position | Special position
Shift position
Only for normal beakers and external positions.
Rinse position
Only for normal beakers and external positions.
Special position
Only for normal beakers and external positions.
Lift rate
Lift rate for manual control on the selected tower.
Input range
5 - 25 mm/s
Default value
25 mm/s
[Start]
Starts moving to the target position. After the start, the button changes to
[Stop], the two lower buttons are shown as inactive (gray) and Move... is
shown as status message instead of Ready.
Moves the lift upwards for as long as this button is pressed.
Moves the lift downwards for as long as this button is pressed.