14 end base point of the baseline, 4 calibration curve results, 1 coefficient k(0) – Metrohm viva 1.0 Manual User Manual

Page 613

background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

viva 1.0

■■■■■■■■

601

Description

Start base point, where the baseline of the peak evaluation begins.

Decimal places

0…3…6

Unit

V

5.6.7.3.14

End base point of the baseline

Variable name

RS.’Command name'.'Substance name'.VAR{x}.REP{y}.END (with
call-up of the VA measuring command by CALL VA)

RS.’Command name'.'Substance name'.VMS.REP{y}.END (with
call-up of the VA measuring command by CALL VMS)

RS.’Command name'.'Substance name'.INTERCEPT.REP{y}.END
(with call-up of the VA measuring command by CALL INTERCEPT)

RS.’Command name'.'Substance name'.'CALL
COND'{x}.REP{y}.END
(with call-up of the VA measuring command by
CALL COND)

Description

End base point, where the peak evaluation baseline ends.

Decimal places

0…3…6

Unit

V

5.6.7.4

Calibration curve results

Standard results of the calibration curve calculation are referred to as cali-
bration curve results
. They are created once per substance.

The results of the regression (CALC0 to CALC4 and R2) are calculated as
soon as a sufficient number of calibration points of different concentra-
tions is available for the selected calibration curve type. The variables will
be recalculated once again and their old value will be overwritten as soon
as another calibration point is added.

5.6.7.4.1

Coefficient k(0)

Variable name

RS.’Command name’.’Substance name’.CALC0

Description

Zero-order coefficient of the calibration curve or standard addition curve.

Decimal places

0…3…6

Unit

#A (for Evaluation quantity = Height)

#C (for Evaluation quantity = Area)