2 meas tmf - general/hardware – Metrohm viva 1.0 Manual User Manual
Page 465
■■■■■■■■■■■■■■■■■■■■■■
5 Method
viva 1.0
■■■■■■■■
453
5.5.2.5.6.2
MEAS TMF - General/Hardware
Tab: Method
▶ MEAS TMF ▶ Properties... ▶ General/Hardware
Command name
Name of the command.
Entry
25 characters
The general parameters for the instrument are defined on this tab.
Device
Device name
Selection of a device from those available in the device table. Only devices
with which the command can be executed are available.
Selection
'Device name' | not defined
Default value
not defined
not defined
In this case, the device name must be assigned by the user at the start
of the method.
Device type
Display or selection of the instrument type. If an instrument is selected in
the Device name field, the Device type field can no longer be edited
and the instrument type belonging to the instrument is displayed.
If the not defined option is selected in the Device name field, all instru-
ment types or groups with which the command can be executed can be
selected, regardless of the instruments present in the device table.
Selection
089 Photometer
Default value
089 Photometer
Sensor
Temperature measurement
Type of temperature measurement
Selection
continuous | off
Default value
continuous
continuous
A temperature sensor must be connected. The temperature will then
be measured continuously.
off
The temperature will not be measured. The temperature entered man-
ually under Temperature on the Measuring parameters tab will be
used.