2 scan - properties – Metrohm viva 1.0 Manual User Manual
Page 538
5.5 Method run subwindow
■■■■■■■■■■■■■■■■■■■■■■
526
■■■■■■■■
viva 1.0
Identification
Description
FIN
Command status; 1 = Command has ended at
least once, 0, invalid (variable not available) =
Command has never ended
TOU
Timeout status: 1 = Max. timeout expired; 0 =
Max. timeout not expired
5.5.2.9.3.2
SCAN - Properties
Dialog window: Method
▶ SCAN ▶ Properties... ▶ SCAN - 'Command name'
Command name
Name of the command.
Entry
25 characters
Device
Device name
Selection of a device from those available in the device table. Only devices
with which the command can be executed are available.
Selection
'Device name' | not defined
Default value
not defined
not defined
In this case, the device name must be assigned by the user at the start
of the method.
Device type
Display or selection of the instrument type. If an instrument is selected in
the Device name field, the Device type field can no longer be edited
and the instrument type belonging to the instrument is displayed.
If the not defined option is selected in the Device name field, all instru-
ment types or groups with which the command can be executed can be
selected, regardless of the instruments present in the device table.
Selection
894 Professional CVS | 846 Dosing Interface |
814 USB Sample Processor | 815 Robotic USB
Sample Processor XL | 858.0010 Professional
Sample Processor | 858.0020 Professional
Sample Processor | 858.0030 Professional
Sample Processor | 919.0020 IC Autosampler
plus
Default value
894 Professional CVS
Remote Box
Selection of a Remote Box connected to the instrument.