3 Workplace

viva 1.0

■■■■■■■■

115

Checking sample data
Check whether the sample data entered corresponds to the method
variables defined and whether it is valid.

Assigning determination ID
A unique determination ID is assigned for the determination.

Reserving instruments (or instrument modules)
The instruments or their function units used by the instrument-depend-
ent commands of the method are reserved for the duration of the
determination, i.e. they can neither be used on other workplaces nor
operated manually.

Increasing sample number
The Sample number is increased by +1.

5 - Main run
The following actions can be carried out during processing of the loaded
method:

Main track processing
After the start of the main track the status changes to BUSY. The com-
mands of the main track and of the tracks requested by it are pro-
cessed in sequence. If the main track requests tracks that are not free,
it will wait until these tracks can be started.

Interrupting and continuing run
An ongoing determination can be interrupted with [Hold] and
resumed with [Cont] at any time. In this case, all active tracks are
interrupted and resumed together.

Canceling a run
An ongoing determination can be canceled at any time with [Stop].
Afterwards, the exit track (if present) is run through and the determina-
tion is ended.

Stop by error
If an error that causes the determination to be interrupted occurs in an
ongoing determination, the error track (if present) is run through and
the determination is ended.

Processing exit track
If the main track and all tracks requested by it are ended, the exit track
(if present) is run through and the determination is ended. The status
changes to READY after the end of the main track.

3.6.3.8

Single determination - Properties

3.6.3.8.1

Properties - Overview

Dialog window: Workplace

Run Single determination Properties Run win-

dow

Properties - Single determination

The properties for the Single determination tab are set on the following
tabs: