5.5 Method run subwindow
■■■■■■■■■■■■■■■■■■■■■■
422
■■■■■■■■
viva 1.0
Cancel determination
The ongoing determination is canceled, after which the next determi-
nation in the series is started (any exit track that may be present is exe-
cuted, any error track that may be present is not executed).
Cancel determination and series
The ongoing determination is canceled, the next determination in the
series will no longer be started (any exit track that may be present is
executed, any error track that may be present is not executed).
Condition
on | off (Default value: off)
If this check box is activated, the defined action will be triggered if the
evaluation of the condition yields the result 1 (= true).
Condition
Expression for the condition(s) that can be created or edited with the for-
mula editor after pressing on
or by double-clicking in the text field
Entry
1,000 characters
Default value
'empty'
Action
Definition of the action that is to be executed if this stop criterion is fulfil-
led.
Selection
Cancel command | Cancel determination | Can-
cel determination and series
Cancel command
The running command CALL COND is canceled, afterwards the next
command is started in the track.
Cancel determination
The ongoing determination is canceled, after which the next determi-
nation in the series is started (any exit track that may be present is exe-
cuted, any error track that may be present is not executed).
Cancel determination and series
The ongoing determination is canceled, the next determination in the
series will no longer be started (any exit track that may be present is
executed, any error track that may be present is not executed).