5 Method

viva 1.0

■■■■■■■■

557

5.6.3.3

Editing recognition

Subwindow section: Method

Evaluation Substances Recogni-

tion

[Edit] New... / Properties... Substance - Recognition

Measuring command

Shows the measuring command for which the substances can be defined.

Substance

Name of the substance that must be unambiguous within the same table
for each measuring command. The same substance name can however
also be used for a different measuring command.

Entry

1 - 50 characters

Default value

New substance

Active

on | off (Default value: on)

If this check box is activated, then this substance will be evaluated; if it is
deactivated, then no evaluation will take place. Parameter data from inac-
tive substances will also continue to be listed in other tables, but have no
effect on the determination.

Characteristic potential

Verification potential for the assignment of a peak to a substance.

Input range

-5.000 - 5.000 V

Default value

'empty' V
The peak potential determined must be within the
range Characteristic potential ± Tolerance so
that the substance is recognized.

Tolerance

Tolerance of the verification potential for the assignment of a peak to a
substance. The peak potential determined must be within the range Char-
acteristic potential
± Tolerance so that the substance is recognized.

Input range

0.001 - 1.000 V

Default value

0.200 V

Min. width

Minimum peak width for the assignment of a peak to a substance. The
calculated peak width must be larger than Min. width in order for the
substance to be recognized.

Input range

0.001 - (Max. width) V

Default value

0.010 V