beautypg.com

6 properties - dark spectrum, Properties - dark spectrum, Dark spectrum – Metrohm viva 1.0 Manual User Manual

Page 785

background image

■■■■■■■■■■■■■■■■■■■■■■

7 Devices

viva 1.0

■■■■■■■■

773

User

Shows the user logged in at the time of the wavelength calibration.

Method

Shows the method with which the wavelength calibration has been car-
ried out.

Selection

Factory settings | "Method name"

Default value

Factory settings

[Reset calibration]

Resets the device to the factory-set calibration. In this process, Dark
spectrum
and Reference spectrum are deleted.

Determination ID

ID of the determination with which the wavelength calibration has been
carried out. With the Factory settings calibration method, ---- is dis-
played.

The determination ID can be copied and pasted into another field.

Calibration coefficients
Coefficients of the regression function with which the CCD pixels are allo-
cated to a wavelength.

c#

Shows the wavelength coefficients c1 - c4.

7.6.6

Properties - Dark spectrum

Tab: Configuration

Devices [Edit] Properties... Properties - 'Device type' -

'Device name'

Dark spectrum

The Dark spectrum tab shows information on the dark spectrum stored
in the device.

If no dark spectrum has been recorded yet or the calibration has been
reset to Factory settings, -- is displayed in the fields.

Date

Shows the date and time when the dark spectrum was recorded.

User

Shows the user logged in during recording of the dark spectrum.

Method

Shows the method with which the dark spectrum has been recorded.