5.5 Method run subwindow
■■■■■■■■■■■■■■■■■■■■■■
480
■■■■■■■■
viva 1.0
Dosing device
Parameters for automatic dosing. This section is displayed only when
Addition = Add with dosing device.
Dosing rate
Rate (volume per time unit) of expulsion. The maximum rate depends on
the cylinder volume of the buret used and on the diameter of the hose at
the selected dosing port.
Input range
0.01 - 166.00 mL/min
Default value
2.00 mL/min
Selection
maximum
CAUTION
The maximum permitted dosing rate of 4 mL/min may not be exceeded
for capillary tubes with a diameter < 1 mm.
Filling rate
Rate (volume per time unit) of filling. The maximum rate depends on the
cylinder volume of the buret used and on the diameter of the hose at the
selected fill port.
Input range
0.01 - 166.00 mL/min
Selection
maximum
Default value
maximum
5.5.2.7.4
ADD SAMPLE
5.5.2.7.4.1
ADD SAMPLE - Overview
Dialog window: Method
▶ ADD AUX ▶ Properties... ▶ ADD SAMPLE - 'Command
name'
Command for the addition of the sample.
Appearance
The command has the following appearance:
Parameters
The parameters for the ADD SAMPLE command are configured in the
following dialog window: