beautypg.com

3 stirring, 1 stirring - overview, Stirring – Metrohm viva 1.0 Manual User Manual

Page 814

background image

8.4 Manual control - Functions

■■■■■■■■■■■■■■■■■■■■■■

802

■■■■■■■■

viva 1.0

Input range

0.01 - 166.00 mL

Default value

4.00 mL

Filling rate

The buret will be filled at this rate after dosing. The maximum filling rate
depends on the cylinder volume of the dosing unit used. If the entered fill-
ing rate is too high for the selected dosing device, it will automatically be
reduced during filling to the largest possible value.

NOTE

The filling rate should be reduced for viscous liquids.

Selection

maximum

Default value

maximum

Input range

0.01 - 166.00 mL

Default value

4.00 mL

[Dosing]

Starts manual dosing for the selected dosing device. Dosing continues as
long as the button is pressed down. The dosed volume is displayed live.

NOTE

Parameters that are modified after dosing has started are not valid until
the next dosing procedure.

[Fill]

Starts filling of the buret for the selected dosing device. The volume dis-
play shows the added volume until the filling procedure has ended.

8.4.3

Stirring

8.4.3.1

Stirring - Overview

Program part: Manual control

If a Stirrer connected via MSB or stirrer connector is selected in the sub-
window for device selection (see Chapter 8.3, page 795), then the func-
tions and parameters that are possible with the stirrers are displayed in the
subwindow for functions/parameters.

Stirrer functions
The stirrer functions are displayed on the following tabs:

Switching on/off

Operating time