7.5 814/815 USB Sample Processor

■■■■■■■■■■■■■■■■■■■■■■

762

■■■■■■■■

viva 1.0

ple positions (see Chapter 6.11.3, page 722) and special beakers (see
Chapter 6.11.3.2, page 722)
are compared with the
Min. beaker
radius
. If this minimum beaker radius is not achieved, the run is stopped
and an error message is output. No check will be made if off is enabled.

Input range

1.0 - 100.0 mm

Selection

off

Default value

off

Lift rate

Lift rate for manual operation.

Input range

3 - 25 mm/s

Default value

25 mm/s

Axial distance

Distance between the axis of rotation of the sample rack and the swing
axis of the robotic arm.

814 USB Sample Processor, 858 Professional Sample Processor
Input range

100.0 - 300.0 mm

Default value

166.0 mm

815 Robotic USB Sample Processor XL, 919 IC Autosampler plus
Input range

100.0 - 300.0 mm

Default value

196.0 mm

Swing Head
This parameter is only shown only if a Swing Head is mounted on the
tower.

Serial number

Shows the serial number of the Swing Head.

Swing position

Lift position that is valid for all four external positions and at which the
robotic arm turns to the external positions.

Input range

0 - 235 mm

Default value

0 mm

Rinse position

Rinse position that is valid for all four external positions and which can be
approached with the LIFT command.

Input range

0 - 235 mm

Default value

0 mm