4 meas opt - evaluations, Evaluations – Metrohm viva 1.0 Manual User Manual
Page 450
![background image](/manuals/586912/450/background.png)
5.5 Method run subwindow
■■■■■■■■■■■■■■■■■■■■■■
438
■■■■■■■■
viva 1.0
Temperature
This section is displayed only when Device type = 089 Photometer.
Temperature
Measuring temperature which can be entered manually. The temperature
is measured continuously when Temperature measurement = contin-
uous (tab General/Hardware).
Input range
20.0 - 60 °C
Default value
25.0 °C
5.5.2.5.4.4
MEAS Opt - Evaluations
Tab: Method
▶ MEAS Opt ▶ Properties... ▶ MEAS Opt - 'Command name' ▶ Eval-
uations
Command name
Name of the command.
Entry
25 characters
The following additional methods for evaluation of measurement curves
can be activated and defined on this tab.
Fixed endpoint evaluation
Fixed endpoint evaluation
on | off (Default value: off)
If this check box is activated, then the associated values for the other
measured quantities will be interpolated with a fixed measured quantity
(Measured value or Time) for the fixed endpoint from the measuring
point list.
Moves the selected line up (changes the sequence).
Moves the selected line down (changes the sequence).
[New]
Opens the fixed endpoint evaluation # dialog window to enter a new
fixed endpoint (see Chapter 5.5.2.5.4.6, page 442).
[Properties]
Opens the fixed endpoint evaluation # dialog window to edit the fixed
endpoint selected in the table (see Chapter 5.5.2.5.4.6, page 442).