5 Method

viva 1.0

■■■■■■■■

543

Type = Number
Input range

-1.0E-99 - 1.0E+99

Type = Text
Entry

max. 100 characters

Type = Date/Time
Selection

'YYYY-MM-DD' or 'YYYY-MM-DD hh:mm:ss'

NOTE

If the sample identifications are still invalid at the start of the method
and are to wait for the REQUEST command before being entered in
the run, then the Check at start option in the START command must
be deactivated for these sample data variables.

Sample type

Sample type.

Selection

Sample | Standard

Default value

Sample

Sample position

Position of the sample on the sample changer rack. This number can be
used to move to the sample position with a MOVE command when Tar-
get position
= Sample position.

Input range

1 - 999

Default value

1

Sample amount

Numerical value for the sample amount for which the standard result is
calculated (see Glossary, page 872).

Entry

Number with 10 digits at most

Default value

1.0

Sample amount unit

Unit of the sample amount. Any text can be entered into this field.

Entry

16 characters

Selection

L | mL | µL | g | mg | µg | pieces

Default value

mL