beautypg.com

8 system variables, System variables – Metrohm viva 1.0 Manual User Manual

Page 43

background image

■■■■■■■■■■■■■■■■■■■■■■

2 General program functions

viva 1.0

■■■■■■■■

31

Available determination variables

Designation

Description

Data type

DUR

Duration of the determination in s

Number

STT

Moment at which the determination has been started

Date/Time

2.3.3.8

System variables

Dialog window: Formula editor

Variables

System variables are general variables which are adopted in the determi-
nation at the start of the determination. They are assigned neither to indi-
vidual commands nor to determinations. The Determination variables
field of the formula editor lists all the System variables that are available
for the current method.

Syntax
'SV.Variable identification'

Examples: 'SV.SIN', 'SV.SLI'

System variables can be selected directly in the formula editor under Vari-
ables

System variables so as to avoid syntax errors.

Available system variables

Designation

Description

Data type

ACC

Autostart actual counter

Number

ACE

Autostart setpoint counter

Number

DID

Determination ID

Text

FUN

Full name of the logged-in user

Text

ORG

Method run: 1 = original determination, 0 = reprocessing

Number

REM

Remarks

Text

RUN

Sample number

Number

SEN

Indication whether the end of the sample table has been reached; 1 = yes, 0 =
no

Number

SIN

Indication whether the determination has been started as a single determination
or within a series; 1 = single determination, 0 = series determination

Number

SLI

Sample table actual line (number)

Number

STC

Start counter

Number

STO

Indication whether the determination has been stopped (manual stop, stop via
SEND command, emergency stop) or terminated normally; 1 = stopped, 0 =
terminated normally

Number

USN

Short name of the logged-in user

Text