8 system variables, System variables – Metrohm viva 1.0 Manual User Manual
Page 43
■■■■■■■■■■■■■■■■■■■■■■
2 General program functions
viva 1.0
■■■■■■■■
31
Available determination variables
Designation
Description
Data type
DUR
Duration of the determination in s
Number
STT
Moment at which the determination has been started
Date/Time
2.3.3.8
System variables
Dialog window: Formula editor
▶ Variables
System variables are general variables which are adopted in the determi-
nation at the start of the determination. They are assigned neither to indi-
vidual commands nor to determinations. The Determination variables
field of the formula editor lists all the System variables that are available
for the current method.
Syntax
'SV.Variable identification'
Examples: 'SV.SIN', 'SV.SLI'
System variables can be selected directly in the formula editor under Vari-
ables
▶ System variables so as to avoid syntax errors.
Available system variables
Designation
Description
Data type
ACC
Autostart actual counter
Number
ACE
Autostart setpoint counter
Number
DID
Determination ID
Text
FUN
Full name of the logged-in user
Text
ORG
Method run: 1 = original determination, 0 = reprocessing
Number
REM
Remarks
Text
RUN
Sample number
Number
SEN
Indication whether the end of the sample table has been reached; 1 = yes, 0 =
no
Number
SIN
Indication whether the determination has been started as a single determination
or within a series; 1 = single determination, 0 = series determination
Number
SLI
Sample table actual line (number)
Number
STC
Start counter
Number
STO
Indication whether the determination has been stopped (manual stop, stop via
SEND command, emergency stop) or terminated normally; 1 = stopped, 0 =
terminated normally
Number
USN
Short name of the logged-in user
Text