beautypg.com

Metrohm viva 1.0 Manual User Manual

Page 689

background image

■■■■■■■■■■■■■■■■■■■■■■

6 Configuration

viva 1.0

■■■■■■■■

677

10 mL

33.0 mL/min

20 mL

66.0 mL/min

50 mL

166.0 mL/min

NOTE

Enter lower rates if you are using capillaries or if the solution has a
higher viscosity.

Tubing parameters
Definition of length and diameter of the tubing connected to the dosing
unit. In addition, the port assignment can be modified.

NOTE

The lengths and diameters must correspond to the actually used tubing,
because this is a prerequisite to ensure a correct execution of the PREP
and EMPTY commands. Please visit the Metrohm Web site (

Connec-

tors made of plastic

) for information on the length and diameter of

Metrohm tubing.

Dosing port 1

Port

Port to be used as dosing port 1.

Selection

Port 1 | Port 2 | Port 3 | Port 4

Default value

Port 1

Length

Length of the tubing on dosing port 1.

Input range

0.0 - 999.9 cm

Default value

80.0 cm

Diameter

Diameter of the tubing on dosing port 1.

Input range

0.0 - 9.9 mm

Default value

0.3 mm

Dosing port 2