5 automatically calculated result variables, Automatically calculated result variables – Metrohm viva 1.0 Manual User Manual
Page 39
■■■■■■■■■■■■■■■■■■■■■■
2 General program functions
viva 1.0
■■■■■■■■
27
Designation
Description
Commands
SAN
Current absolute swing angle of the robotic arm in ° (entry upon exiting the com-
mand)
MOVE, SWING
SDV
Actual value of the standard deviation when the command is canceled
CALL COND
SPO
Current external position (entry when ending the command); 0 means invalid
position
SWING
STC{x}
Concentration of the standard x (1 - 50) used for the command
CAL LOOP Opt
STY
Type of stop with which the command was stopped: 1 = normal; 0 = manual
or after error
CAL MEAS
Opt, CAL
Spec, CVS,
MEAS Opt,
MEAS Opt
Conc, MEAS
Ref, MEAS
Spec, MEAS
TMF
TMF
Transmission factor (dimensionless)
MEAS TMF
TOL
Tolerance within which a measured peak can be assigned to a calibration wave-
length.
CAL Spec
TOU
Timeout status: 1 = Max. waiting time expired; 0 = Max. waiting time not
expired
RECEIVE,
SCAN, TRANS-
FER
VOL
Dosed volume increment in L
ADD AUX,
ADD SAMPLE,
ADD SAMPLE
DT, ADD STD,
LQH
VOLTOT
Dosed total volume in L
ADD SAMPLE
DT, ADD STD
2.3.3.5
Automatically calculated result variables
Dialog window: Formula editor
▶ Variables
The automatically calculated result variables are automatically generated
by the program for the VA measuring commands contained in the
method during the evaluation. The Variables field of the formula editor
lists all the variables which are available for the current method.
Syntax
'RS.Command name.Substance name.VAR{x}.REP{x}.Variable
identification'
Examples: 'RS.CVS_1.Cu.VAR{2}.REP{3}.HGT
You can select the method variables directly in the formula editor under
Variables
▶ Result variables in order to avoid syntax errors.