2 cal loop opt - properties – Metrohm viva 1.0 Manual User Manual
Page 476
5.5 Method run subwindow
■■■■■■■■■■■■■■■■■■■■■■
464
■■■■■■■■
viva 1.0
Command variables
The following command variables are generated by the CAL LOOP Opt
command in the method run and can be used in formulas under the des-
ignation 'Command name.Variable identification':
Identification
Description
BSY
Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY; invalid (variable not available) = Command
has never been started
C0
Coefficient c0 calculated from the calibration
C1
Coefficient c1 calculated from the calibration
C2
Coefficient c2 calculated from the calibration
C3
Coefficient c3 calculated from the calibration
COD
Coefficient of determination r
2
), which is calculated
on the basis of the calibration function and of the
confidence interval
FIN
Command status; 1 = Command has ended at
least once, 0, invalid (variable not available) =
Command has never ended
LCO
Loop counter = current number of completed runs,
both of Repeat loops and While loops
LST
Start time of the loop command (Date/Time)
OUT
Number of outliers
STC{x}
Concentration of the standard x (1 - 50) used for the
command
5.5.2.6.3.2
CAL LOOP Opt - Properties
Dialog window: Method
▶ CAL LOOP Opt ▶ Properties... ▶ CAL LOOP Opt - 'Com-
mand name'
Command name
Name of the command.
Entry
25 characters
Standards
Unit conc.
Selection of the concentration unit for the standards.