5.5 Method run subwindow
■■■■■■■■■■■■■■■■■■■■■■
470
■■■■■■■■
viva 1.0
Parameters for the Avantes spectrometer
Wavelength
Wavelength at which the measured quantity (absorbance) is recorded
Input range
100.0 - 2,000.0 nm
Default value
400.0 nm
Blank value
Value by which the measured quantity (absorbance) is corrected. Here an
expression can also be entered through the formula editor (e.g. xx.EME
from other measuring command) by which every measured absorbance
value can then be corrected.
Input range
-1E+12 - 1E+12 (max. 10 digits) mAU
Default value
0 mAU
Integration time
Integration time for absorbance measurement.
Input range
0.01 - 600,000 ms
Default value
6 ms
Averaged spectra
The number of spectra that are acquired and averaged.
Input range
1 - 10,000
Default value
10
Smoothing
Number of neighboring pixels with which the value for each measurement
pixel is determined.
Input range
0 - 100 pixels
Default value
0 pixels
Parameters for the 089 Photometer
Blank value
Value by which the measured quantity (absorbance) is corrected. Here an
expression can also be entered through the formula editor (e.g. xx.EME
from other measuring command) by which every measured absorbance
value can then be corrected.
Input range
-4,000 - 4,000 mAU
Default value
0 mAU