5 Method
viva 1.0
■■■■■■■■
503
Number
Specification of the number of the rack position. This parameter is visible
only when Move = Rack position or Special beaker. The position can,
however, also be entered as a formula. The formula editor is opened with
a right click on the input field.
Move = Rack position
Input range
1 - 999
Default value
1
Move = Special beaker
Input range
1 - 16
Default value
1
Move angle
This parameter specifies the angle by which the sample rack is to be rota-
ted relative to its momentary position. It is visible only when Move = Rel-
ative angle.
814, 815, 858
Input range
-359.90 - 359.90 °
Default value
5.00 °
Beaker test
Selection of the action to be carried out when the beaker sensor defined
for the rack detects no beaker at the target position. If the robotic arm is
selected as beaker sensor, then a Swing Head must be installed. A work
position that is also defined in the rack data must be defined in such a
way that a contact exists between beaker sensor and beaker. This work
position will be automatically moved to for beaker recognition after the
command MOVE. This section is not visible when Move = Relative
angle.
Selection
Display message | Cancel determination | Can-
cel determination and series
Default value
Display message
Display message
A message is displayed, the determination is canceled and the next
determination of the series is started after confirmation of the mes-
sage. If an exit track is present, it will be called. No error track will be
called.