5 Method
viva 1.0
■■■■■■■■
431
Identification
Description
BP{#}.TIM
Time for the break point # (1 - 9) in s
BP{#}.TRN
Transmission for the break point x (1 - 9) in %
BSY
Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY; invalid (variable not available) = Command
has never been started
DBL
Total duration for the processing of the command in
s
EME
End measured value (measured value after processing
of the command) in the unit of the measured value
ETE
1)
End temperature (temperature following processing
of the command) in °C
FIN
Command status; 1 = Command has ended at
least once, 0, invalid (variable not available) =
Command has never ended
FP{#}.[MEA]
Measured value for the fixed endpoint # (1 - 9) in the
unit of the measured value
FP{#}.CNT
Intensity for the fixed endpoint x (1 - 9) in counts
FP{#}.TEM
1)
Temperature for the fixed endpoint # (1 - 9) in °C
FP{#}.TIM
Time for fixed endpoint # (1 - 9) in s
FP{#}.TRN
Transmission for the fixed endpoint x (1 - 9) in %
IME
Initial measured value (measured value before start
conditions are processed) in the unit of the measured
value
ITE
1)
Initial temperature (temperature before start condi-
tions are processed) in °C
LP.CAx
Calculated value × (1 - 3) for the last measuring point
on the measuring point list
LP.CNT
Transmission of the last measuring point of the mea-
suring point list in %
LP.EXx
External value × (1 - 3) for the last measuring point of
the measuring point list
LP.MEA
Measured value for the last measuring point of the
measuring point list in the unit of the measured value
LP.TEM
1)
Temperature for the last measuring point of the mea-
suring point list in °C