7.5 814/815 USB Sample Processor

■■■■■■■■■■■■■■■■■■■■■■

768

■■■■■■■■

viva 1.0

Acoustic signal

on | off (Default value: off)

If this check box is activated, an acoustic signal will be emitted in addition
to the message.

Action
The actions can only be edited if monitoring is switched on.

If it is found during monitoring that the validity period has expired, then
one of the following actions will be triggered automatically during the
start test:

Selection

Record message | Display message | Cancel
determination

Default value

Display message

Record message
The message that the validity period has expired will be automatically
saved with the determination.
Display message
A message is displayed and you can select whether you want to con-
tinue the run or cancel it. If the run is continued then the message that
the validity period has expired will be automatically saved with the
determination.
Cancel determination
The determination is not started. The following message must be con-
firmed with [OK].

7.5.8

Robotic arm configuration

Dialog window: Configuration

Devices [Edit] Properties... Properties - 'De-

vice type' - 'Device name'

Tower [Configuration] Robotic arm configura-

tion

The specific settings for the robotic arm mounted on a Swing Head can be
made in the Robotic arm configuration dialog window.

Swing offset

Swing angle offset for specific robotic arm model. This value must be set
according to the leaflet supplied with the robotic arm.

Input range

-270.0 - 270.0 °

Default value

0.0 °

Max. swing range

Maximum usable swing range for specific robotic arm model. Each robotic
arm model displays a different value on the basis of its construction. This