6 Configuration

viva 1.0

■■■■■■■■

697

Action
If during calibration data monitoring it is found that its validity has
expired, then one of the following actions will be triggered automatically
during the start test:

Editable only if Calibration data monitoring = on, otherwise inactive.

Selection

Record message | Display message | Cancel
determination

Default value

Display message

Record message
The message that the validity of the calibration has expired will be
automatically saved with the determination.
Display message
A message is displayed and you can select whether the run is to be
continued or canceled. If the run is continued then the message that
the validity of the calibration has expired will be saved automatically
with the determination.
Cancel determination
The ongoing determination will be automatically canceled. The follow-
ing message must be confirmed with [OK].

6.8.3.4

Properties - Limit values

Tab: Configuration

Colorimetric sensors Edit Properties... Limit values

Monitoring number of outliers

on | off (Default value: off)

If this check box is activated, then the number of outliers during a new
calibration is monitored. If this number exceeds the defined limit value,
then a corresponding message will appear and you can decide whether
the calibration should nevertheless be applied or if the previous calibration
data should be retained.

Upper limit

Maximum number of permitted outliers.

Only editable for Monitoring number of outliers = on, otherwise inac-
tive.

Input range

0 - 9

Default value

1